Calibration of UV-Radiation Dose with Irradiation Time for CR-39 Track Detector by Using UV-visible Spectrophotometer

Received: 10 June 2015     Accepted: 30 June 2015     Published: 1 July 2015
Views:       Downloads:
Abstract

Nuclear track detector type CR-39 irradiated by two UV-irradiation methods at wavelength 254nm . First method was PEN-ray-lamp UV-irradiation with the distance 4cm between UV source to CR-39 detector and the range of irradiation time - TD were 10,100 and 1000 s . Second method was FLX-20 transiluminator 15W energy amount with the distance 10 cm between UV source to CR-39 detector and therange of irradiation dose - DT were 2 , 6 , 9 and 10J/cm2. The absorbance-A was determine from UV-visible spectroscopy at the measuring range from 300 – 500 nm for each method .The absorbance differences [ΔA]350 and [ΔA]550 for irradiated and un-irradiated CR-39 detector were measured between absorbance value at wavelength 350nm and 550nm respectively . In this study determined the linear relationship between the increasing of TD(s) and DT(J/cm2) with increasing of [ΔA]350 and [ΔA]550 .This work obtained that FLX-20transiluminator ( second method) was used as a calibration route to PEN-ray-lamp ( first method) after measuring the absorbance deference [ΔA]350 and [ΔA] 550 for CR-39 detector.

Published in American Journal of Modern Energy (Volume 1, Issue 2)
DOI 10.11648/j.ajme.20150102.11
Page(s) 25-29
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2015. Published by Science Publishing Group

Keywords

Solid-State Detectors, Visible and Ultraviolet Spectra, Dosimetry, Ultraviolet Radiation Effects

References
[1] Chun ChunT. , " Investigations of the Effects of UV Irradiation on the Etching Behavior of CR-39 Solid State Nuclear Track Detector " Department of Physics and Materials Science, Ms. Theses , March( 2007) .
[2] Nagpal, J. S., " Ultraviolet dosimetry using thermoluminescent phosphors " Radiation Standards and Instrumentation Div., Bhabha Atomic Research Centre , Mumbai (India) ; Bhabha Atomic Research Centre, Mumbai (India) , T-24-5 , P-8-88(1998)
[3] TseK.C.C. , F. M. F. Ng and YuK . N." Photo-degradation of PADC by UV radiation at various wavelengths "Polymer Degradation and Stability , 91, 2380-2388, (2006)
[4] Hussain A. Al-Jobouri , Nasreen R . Jber , Abbas H. Al-Shukrawi and Mazin K. Hamid " Physiochemical properties of crystalline etch products for CR-39 track detector after α-particles irradiation " Advances in Applied Science Research , 4(4) , 501-507 (2013)
[5] Abdullah Fathi Aly Saad " Changes in the thermal properties ofPADC film - based NTDs produced by high doses of γ-radiation " Nuclear Instruments and Methods in Physics Research B , Volume: B 325 , (2014) 79–83
[6] Tse, K. C. C. , Nikezic, D. and Yu, K. N." Effects of UVC irradiation on alpha-particle track parameters in CR-39 " Radiation Measurements , 43 , S98–S101(2008)
[7] Hareesh , K . , Pandey , A . K . , Sangappa , Y. , Ravishankar Bhat , Venkataraman , A. and Sanjeev, G. " Changes in the properties of Lexan polycarbonate by UV irradiation " Nuclear Instruments and Methods in Nuclear Instruments and Methods in Physics Research B , Physics Research, B 295 , 61-68 ( 2013 )
[8] KnuschkeP Barth J. "Biologicallyweightedpersonal UV dosime try " J . Photochem. Photobiol. B. , 36(1), 77-83 ( 996).
[9] Shweikani , R. , Raja , G. and Sawaf A. A. " The possibility of using plastic detectors CR-39 as UV Dosimeters " Radiation Measurements , 35 , 281 - 285(2002).
[10] Seckmeyer , G. , Klingebiel , M . , Riechelmann , S. , LohseI. and McKenzie , R. L. , Ben Liley, J. " A critical assessment of two types of personal UV dosimeters " Photochemistry and Photobiology , 88(1) 215-22 , ( 2011) .
[11] Hareesh, K. , Sanjeev G. ,PandeyA. K. and Vijayalakshmi , Rao. "Characterization of UV-irradiated Lexan polycarbonate films" Iran Polym J . 22 , 341–349(2013)
[12] Abu - Jarad , F. , El Hadidy , M . and Al – Jarallah , MI. " The Use Of Nuclear Track Detectors As An Environmental Solar Ultraviolet Radiation Dosimeter" Pergamon - Elsevier science ltd . , Radiation Measurements , 28 ,409-412( 1997 )
Cite This Article
  • APA Style

    Hussain A. Al-Jobouri, Mohamed Bechir Ben Hamida, Nada F. Tawfiq, Fala Hatem Taha. (2015). Calibration of UV-Radiation Dose with Irradiation Time for CR-39 Track Detector by Using UV-visible Spectrophotometer. American Journal of Modern Energy, 1(2), 25-29. https://doi.org/10.11648/j.ajme.20150102.11

    Copy | Download

    ACS Style

    Hussain A. Al-Jobouri; Mohamed Bechir Ben Hamida; Nada F. Tawfiq; Fala Hatem Taha. Calibration of UV-Radiation Dose with Irradiation Time for CR-39 Track Detector by Using UV-visible Spectrophotometer. Am. J. Mod. Energy 2015, 1(2), 25-29. doi: 10.11648/j.ajme.20150102.11

    Copy | Download

    AMA Style

    Hussain A. Al-Jobouri, Mohamed Bechir Ben Hamida, Nada F. Tawfiq, Fala Hatem Taha. Calibration of UV-Radiation Dose with Irradiation Time for CR-39 Track Detector by Using UV-visible Spectrophotometer. Am J Mod Energy. 2015;1(2):25-29. doi: 10.11648/j.ajme.20150102.11

    Copy | Download

  • @article{10.11648/j.ajme.20150102.11,
      author = {Hussain A. Al-Jobouri and Mohamed Bechir Ben Hamida and Nada F. Tawfiq and Fala Hatem Taha},
      title = {Calibration of UV-Radiation Dose with Irradiation Time for CR-39 Track Detector by Using UV-visible Spectrophotometer},
      journal = {American Journal of Modern Energy},
      volume = {1},
      number = {2},
      pages = {25-29},
      doi = {10.11648/j.ajme.20150102.11},
      url = {https://doi.org/10.11648/j.ajme.20150102.11},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.ajme.20150102.11},
      abstract = {Nuclear track detector type CR-39 irradiated by two UV-irradiation methods at wavelength 254nm . First method was PEN-ray-lamp UV-irradiation with the distance 4cm between UV source to CR-39 detector and the range of irradiation time - TD were 10,100 and 1000 s . Second method was FLX-20 transiluminator 15W energy amount with the distance 10 cm between UV source to CR-39 detector and therange of irradiation dose - DT were 2 , 6 , 9 and 10J/cm2. The absorbance-A was determine from UV-visible spectroscopy at the measuring range from 300 – 500 nm for each method .The absorbance differences [ΔA]350 and [ΔA]550 for irradiated and un-irradiated CR-39 detector were measured between absorbance value at wavelength 350nm and 550nm respectively . In this study determined the linear relationship between the increasing of TD(s) and DT(J/cm2) with increasing of [ΔA]350 and [ΔA]550 .This work obtained that FLX-20transiluminator ( second method) was used as a calibration route to PEN-ray-lamp ( first method) after measuring the absorbance deference [ΔA]350 and [ΔA] 550 for CR-39 detector.},
     year = {2015}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Calibration of UV-Radiation Dose with Irradiation Time for CR-39 Track Detector by Using UV-visible Spectrophotometer
    AU  - Hussain A. Al-Jobouri
    AU  - Mohamed Bechir Ben Hamida
    AU  - Nada F. Tawfiq
    AU  - Fala Hatem Taha
    Y1  - 2015/07/01
    PY  - 2015
    N1  - https://doi.org/10.11648/j.ajme.20150102.11
    DO  - 10.11648/j.ajme.20150102.11
    T2  - American Journal of Modern Energy
    JF  - American Journal of Modern Energy
    JO  - American Journal of Modern Energy
    SP  - 25
    EP  - 29
    PB  - Science Publishing Group
    SN  - 2575-3797
    UR  - https://doi.org/10.11648/j.ajme.20150102.11
    AB  - Nuclear track detector type CR-39 irradiated by two UV-irradiation methods at wavelength 254nm . First method was PEN-ray-lamp UV-irradiation with the distance 4cm between UV source to CR-39 detector and the range of irradiation time - TD were 10,100 and 1000 s . Second method was FLX-20 transiluminator 15W energy amount with the distance 10 cm between UV source to CR-39 detector and therange of irradiation dose - DT were 2 , 6 , 9 and 10J/cm2. The absorbance-A was determine from UV-visible spectroscopy at the measuring range from 300 – 500 nm for each method .The absorbance differences [ΔA]350 and [ΔA]550 for irradiated and un-irradiated CR-39 detector were measured between absorbance value at wavelength 350nm and 550nm respectively . In this study determined the linear relationship between the increasing of TD(s) and DT(J/cm2) with increasing of [ΔA]350 and [ΔA]550 .This work obtained that FLX-20transiluminator ( second method) was used as a calibration route to PEN-ray-lamp ( first method) after measuring the absorbance deference [ΔA]350 and [ΔA] 550 for CR-39 detector.
    VL  - 1
    IS  - 2
    ER  - 

    Copy | Download

Author Information
  • Department of Physics College of Science, AL-Nahrain University, Baghdad, Iraq

  • High School of Sciences and Technology of Hammam Sousse (ESSTHS), Department of Physics, University of Sousse, Sousse, Tunisia

  • Department of Physics College of Science, AL-Nahrain University, Baghdad, Iraq

  • Department of Physics College of Science, AL-Nahrain University, Baghdad, Iraq

  • Sections